Published March 2008 | Version Published
Journal Article Open

Image distortion in thermoacoustic tomography caused by microwave diffraction

  • 1. ROR icon Washington University in St. Louis

Abstract

We report an intrinsic image distortion in microwave-induced thermoacoustic tomography. The distortion, due to microwave diffraction in the object to be imaged, leads to nonuniform excitation of acoustic pressure during microwave illumination. Both numerical simulations and phantom experiments demonstrate this phenomenon. A method of partial correction is also provided.

Additional Information

© 2008 American Physical Society. Received 18 October 2007; revised manuscript received 11 January 2008; published 28 March 2008. This project was sponsored in part by the National Institute of Health Grants No. R01 NS46214 and R01 EB000712.

Attached Files

Published - PhysRevE.77.031923.pdf

Files

PhysRevE.77.031923.pdf

Files (303.1 kB)

Name Size
md5:7631ca374751676e9a374d8569b6ea70
303.1 kB Preview Download

Additional details

Identifiers

Eprint ID
70562
Resolver ID
CaltechAUTHORS:20160923-134112679

Funding

NIH
R01 NS46214
NIH
R01 EB000712

Dates

Created
2016-09-28
Created from EPrint's datestamp field
Updated
2021-11-11
Created from EPrint's last_modified field