Published September 2013 | Version public
Book Section - Chapter

Capacitive proximity communication with distributed alignment sensing for origami biomedical implants

  • 1. ROR icon California Institute of Technology

Abstract

Origami implant design is a 3D integration technique which addresses size and cost constraints in biomedical implants. A capacitive proximity interconnect that enables this technique is presented. The interconnect embeds an alignment sensor that measures link quality directly and simplifies its adaptation to alignment. The sensor and transceiver share functional blocks, saving power and area. Data rates from 10-60 Mbps are achieved over 4-12 μm of parylene-C, with efficiencies up to 0.180 pJ/bit.

Additional Information

© 2013 IEEE. The authors thank M. Monge for helpful technical discussions, K. Potter for advice on the test setup, J. Park and Y. Liu for parylene fabrication, STMicroelectronics for chip fabrication and the NSF for funding support.

Additional details

Identifiers

Eprint ID
92866
DOI
10.1109/CICC.2013.6658445
Resolver ID
CaltechAUTHORS:20190213-074612379

Funding

NSF

Dates

Created
2019-02-13
Created from EPrint's datestamp field
Updated
2023-03-15
Created from EPrint's last_modified field