We introduce a novel method to correct for imperfect indenter geometry and frame compliance in instrumented indentation testing with a spherical indenter. Effective radii were measured directly from residual indentation marks at various contact depths (ratio of contact depth to indenter radius between 0.1 and 0.9) and were determined as a function of contact depth. Frame compliance was found to depend on contact depth especially at small indentation depths, which is successfully explained using the concept of an extended frame boundary. Improved representative stress-strain values as well as hardness and elastic modulus were obtained over the entire contact depth.

Additional Information

© 2009, Materials Research Society. (Received 31 January 2009; accepted 22 June 2009) This research was supported in part by the Seoul R&BD Program (Grant No. TR080564) of the Seoul Development Institute, Korea, and in part by the Center for Nanoscale Mechatronics and Manufacturing under The 21st Century Frontier R&D Program (Grant No. M102KN010006-08K1401-00621) of the Ministry of Education, Science and Technology, Korea.

Attached Files

Published - Kang2009p5919J_Mater_Res.pdf

Files

Kang2009p5919J_Mater_Res.pdf

Files (1.3 MB)

Name Size
md5:95bd379e85965d193213e6b0a8f10915
1.3 MB Preview Download

Additional details

Identifiers

Eprint ID
16021
Resolver ID
CaltechAUTHORS:20090923-143135393

Funding

Seoul Development Institute
TR080564
Ministry of Education, Science and Technology (Korea)
M102KN010006-08K1401-00621

Dates

Created
2009-10-05
Created from EPrint's datestamp field
Updated
2021-11-08
Created from EPrint's last_modified field