Published June 2003 | Version Published
Book Section - Chapter Open

A parylene MEMS flow sensing array

  • 1. ROR icon California Institute of Technology

Abstract

We have demonstrated the first parylene MEMS thermal sensor array capable of detecting small flows down to 0.5 µl/min. The device is fabricated using low-temperature parylene/platinum technology and consists of metallic sensors suspended on a membrane over a bulk micromachined silicon channel. For the first time, insightful results from three different methods of flow sensing using a single device are studied and compared. Multi-mode testing using hot film, calorimetric, and time-of-flight techniques at low overheat ratios has been performed. Thus, this device is biocompatible in both construction and operation.

Additional Information

© 2003 IEEE. This work was funded in part by the Engineering Research Centers Program of the National Science Foundation under Award Number EEC-9402726. We would like to thank Mr Trevor Roper for assistance in fabrication and Mr. Tuan Hoang for help with experiments and proofreading.

Attached Files

Published - 01215566.pdf

Files

01215566.pdf

Files (351.9 kB)

Name Size
md5:6617674c40c266616de9a6b4cdb47a5c
351.9 kB Preview Download

Additional details

Identifiers

Eprint ID
94149
Resolver ID
CaltechAUTHORS:20190326-112949514

Funding

NSF
EEC-9402726

Dates

Created
2019-03-26
Created from EPrint's datestamp field
Updated
2021-11-16
Created from EPrint's last_modified field