Published October 15, 2007 | Version public
Journal Article Open

Electrical microfluidic pressure gauge for elastomer microelectromechanical systems

  • 1. ROR icon University of Southern California
  • 2. ROR icon California Institute of Technology

Abstract

We report on an electrical microfluidic pressure gauge. A polydimethylsiloxane microvalve closes at a characteristic applied pressure determined by the material's properties and the valve's dimensions. Hence, when the same pressure is applied to all valves of a heterogeneous valve array, some valves close while others remain open. The state of the array is combined with knowledge of the respective characteristic closing pressures of the individual valves to yield an estimate of the applied pressure. The state of each valve is obtained by electrical measurements, since the electrical resistance of the respective underlying fluid-filled channel increases by at least two orders of magnitude as the valve closes and its insulating elastomer material interrupts the electrical circuit. The overall system functions as a pressure gauge with electrical readout. This device would be a critical component in active pressure-regulation loops in future integrated microfluidic systems.

Additional Information

©2007 American Institute of Physics. (Received 3 May 2007; accepted 1 September 2007; published online 26 October 2007) The authors thank Alejandra Torres, Christina Morales, and Ali Ghaffari of the Caltech Micro/Nano-Fluidic Foundry for their help with device fabrication. Financial support for this work was provided by NIH Grant Nos. 1RO1 HG002644-01A1 and 1 K99EB007151-01.

Files

KARjap07b.pdf

Files (182.3 kB)

Name Size
md5:375eb0861f68c0036aba1e041324a1d1
182.3 kB Preview Download

Additional details

Identifiers

Eprint ID
9095
Resolver ID
CaltechAUTHORS:KARjap07b

Dates

Created
2007-10-26
Created from EPrint's datestamp field
Updated
2021-11-08
Created from EPrint's last_modified field

Caltech Custom Metadata

Caltech groups
Physics Department