Published July 7, 2008 | Version Accepted Version + Published
Journal Article Open

In vivo burn imaging using Mueller optical coherence tomography

Abstract

We report on the use of a high-speed, fiber-based Mueller-matrix optical coherence tomography system with continuous source-polarization modulation for in vivo burn depth evaluation and healing monitoring. A homemade hand-held probe with integrated optical scanning and beam delivering optics was coupled in the sample arm. In vivo burn imaging was demonstrated on porcine skin at different stages of the wound healing process, where porcine skin was used because of its similarity to the human skin. Thermally damaged region was clearly localized in the depth-resolved phase retardation image extracted from the measured Jones matrix. The burn areas in the OCT images agreed well with the histology. By using a decomposition algorithm developed by our group, we also mapped the local birefringence of the skin. The experimental results demonstrate the system's potential for in vivo burn-depth determination.

Additional Information

© 2008 Optical Society of America. Received 7 Mar 2008; revised 16 May 2008; accepted 16 May 2008; published 26 Jun 2008. This project was sponsored in part by the Department of the Army (Cooperative Agreement Number: DAMD17-97-2-7016) and the National Institutes of Health (R01 CA092415). The content of the information presented in this paper does not necessarily reflect the position or the policy of the government or NMTB. No official endorsement should be inferred.

Attached Files

Published - oe-16-14-10279.pdf

Accepted Version - nihms-1067736.pdf

Files

nihms-1067736.pdf

Files (1.2 MB)

Name Size
md5:54ab4a9f9fd0c8ac198909f8e1b6a915
933.2 kB Preview Download
md5:ac6385d8431fc45381dc3d8bca850e62
305.7 kB Preview Download

Additional details

Identifiers

PMCID
PMC6986309
Eprint ID
70563
Resolver ID
CaltechAUTHORS:20160923-135417190

Funding

Department of the Army
DAMD17-97-2-7016
NIH
R01 CA092415

Dates

Created
2016-09-28
Created from EPrint's datestamp field
Updated
2021-11-11
Created from EPrint's last_modified field