Photoacoustic microscopy with an enhanced axial resolution of 5.8 μm

  • 1. ROR icon Washington University in St. Louis

Abstract

The axial resolution of photoacoustic microscopy (PAM) can be enhanced by reducing the speed of sound within the imaging region of interest. This principle was demonstrated on a previously-reported PAM system, which utilized a 125 MHz ultrasonic transducer for signal detection and the Wiener deconvolution for signal processing. With sound slowed by silicone oil immersion, we have achieved a finest axial resolution of 5.8 μm for PAM, as validated by phantom experiments. The axial resolution was also enhanced in vivo when mouse ears injected with silicone oil were imaged. After injection of silicone oil, the blood vessels were resolved more clearly. When tissue-compatible low-speed liquids become available, this approach may find applications in PAM as well as in other imaging modalities, such as photoacoustic computed tomography and ultrasound imaging.

Additional Information

© 2014 Society of Photo-Optical Instrumentation Engineers. This work was sponsored in part by National Institutes of Health grants DP1 EB016986 (NIH Director's Pioneer Award), R01 CA134539, U54 CA136398, R01 CA157277, and R01 CA159959. L.W. has a financial interest in Microphotoacoustics, Inc. and Endra, Inc., which, however, did not support this work.

Attached Files

Published - 894316.pdf

Files

894316.pdf

Files (469.1 kB)

Name Size
md5:11b6596dea9d2559a24eb16f5b1c4b82
469.1 kB Preview Download

Additional details

Identifiers

Eprint ID
89565
Resolver ID
CaltechAUTHORS:20180912-095829705

Funding

NIH
DP1 EB016986
NIH
R01 CA134539
NIH
U54 CA136398
NIH
R01 CA157277
NIH
R01 CA159959

Dates

Created
2018-09-12
Created from EPrint's datestamp field
Updated
2021-11-16
Created from EPrint's last_modified field

Caltech Custom Metadata

Series Name
Proceedings of SPIE
Series Volume or Issue Number
8943