Published February 1, 1949
| Version public
Journal Article
Limitations of the Use of Potassium Hydroxide Potassium Cyanide Mixtures in Manometric Studies
Creators
Abstract
The manometric study of biological oxidations by the "direct method" of Warburg, when carried out in the presence of cyanide, involves the use of KOH-KCN mixtures (Krebs (4)) for the absorption of carbon dioxide. The following study describes the rate of distillation of HCN from the experimental solution under various conditions, and suggests a remedy for the difficulties occasionally encountered (Machlis (2)) with the above mixtures.
Additional Information
© 1949 American Society for Biochemistry and Molecular Biology.Additional details
Identifiers
- Eprint ID
- 42960
- Resolver ID
- CaltechAUTHORS:20131211-133555079
Dates
- Created
-
2013-12-23Created from EPrint's datestamp field
- Updated
-
2019-10-03Created from EPrint's last_modified field